• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 24 IN. TOUCH SCREEN; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 24 IN. TOUCH SCREEN; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10003
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/03/2021
Event Type  malfunction  
Event Description
It was reported that during a set up for a cori assisted tka surgery, the real intelligence 24 in.Touch screen blacked out right before calibration and turned completely off and did not come back on remainder of case.They did notice it said in small letters in middle of screen ¿no signal detected¿ right before it turned completely off.The procedure was completed, without delay, using the tablet.Patient was not harmed as consequence of this problem.After case they tried to unplug monitor cord on back that would shut it off and plugged it back in, the monitor would show green light on bottom right hand corner of screen but would immediately turn orange and it would say ¿no signal detected¿ in middle of screen, which made them believe that the cord works fine and that it is the monitor that is defected.
 
Manufacturer Narrative
H10: internal complaint reference (b)(4).Section: h3, h6: the product, ri 24 in touch screen us, rob10003, s/n (b)(6) intended for use in treatment was not returned for evaluation; thus, a visual and functional evaluation could not be performed, and a relationship between the reported event and the device could not be determined.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, a factor that may have contributed to the reported symptom may be associated with a component/communication failure.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE 24 IN. TOUCH SCREEN
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12329246
MDR Text Key266864783
Report Number3010266064-2021-00608
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556756256
UDI-Public00885556756256
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10003
Device Catalogue NumberROB10003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/03/2021
Initial Date FDA Received08/16/2021
Supplement Dates Manufacturer Received11/15/2021
Supplement Dates FDA Received11/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-