This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, the cause is likely bending stress was concentrated at deflected position of the coil, shifting the coil pipe, causing the a-wire to get caught.Olympus will continue to monitor the field performance of this device.
|