Model Number URF-P6
Device Problem Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The event date is (b)(6) 2021 when the broken angulation wire was noted.Further inspection of the returned scope found the flexible tip shear and a leak in the biopsy channel.The cause of the broken angulation wire cannot be determined at this time.An investigation is ongoing to obtain additional information regarding the reported event.If additional information is received this report will be supplemented accordingly.
 
Event Description
The customer reported that the scope would not deflect.There was no patient injury reported.The scope was returned to the service center for evaluation and found the angulation wire broken which is a reportable malfunction.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, the cause is likely bending stress was concentrated at deflected position of the coil, shifting the coil pipe, causing the a-wire to get caught.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12333866
MDR Text Key267037198
Report Number8010047-2021-10363
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340802
UDI-Public04953170340802
Combination Product (y/n)N
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/23/2021
Initial Date FDA Received08/17/2021
Supplement Dates Manufacturer Received09/17/2021
Supplement Dates FDA Received09/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-