STRYKER ENDOSCOPY-SAN JOSE PKG., ASSY., STRYKEFLOW 2 WITH TIP; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
|
Back to Search Results |
|
Catalog Number 0250070520 |
Device Problems
Sparking (2595); Insufficient Information (3190)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that there was a thermal event.
|
|
Manufacturer Narrative
|
Alleged failure: midway through the case, the surgical staff noticed a burning smell and thought it was coming from the surgical suction.They replaced the neptune in the room but noticed the smell was coming from the suction irrigator unit when changing out the fluid.Salesforce case number (b)(4).The failure(s) identified in the investigation is consistent with the complaint record.The root cause was fluid ingress causing a short inside of the motor.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that there was a thermal event.
|
|