Model Number BI70002000
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2021
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: bi71000176, serial/lot #: (b)(4).A manufacturer representative went to the site to test the equipment.It was found that the power enclosure board failed and it was replaced.The imaging system then passed the system checkout and was found to be fully functional.The power enclosure was returned to the manufacturer for analysis.It was found that there was a power converter failure.The power conversion enclosure failed bench testing and transistors were shorted.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system used outside of a procedure.It was reported that there was a power conversion enclosure failure.It was noted that the power conversion would try to boot up when the umbilical was disconnected.Otherwise, the system was functioning as designed.No patient was present when the issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12334735
MDR Text Key267035986
Report Number3004785967-2021-01017
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200074
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/05/2021
Initial Date FDA Received08/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse