MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9736242 |
Device Problems
Computer Software Problem (1112); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: 9 736217, serial/lot #: (b)(4).Product id: 9736226, serial/lot #: (b)(4).Note - the system was checked out and the computer was replaced under a separate event.Software analysis was completed and the issue was consistent with a known software issue that is tracked through a medtronic database.(b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system used outside of a procedure.It was reported that when booting on the system, the splash screen would remain for a minimum of 10 minutes and it would not proceed into the software.After restarting the system, the issue was resolved and the manufacturer representative was able to log into the system.No patient present when the issue was identified.It was reported that the system had since checked out as expected.It was reported that the computer had failed and was replaced.
|
|