Model Number 102096-67A
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On (b)(6) 2021, senseonics was made aware of an incident where user's sensor was not detected by neither the old transmitter nor the replacement leading to early sensor removal.
 
Manufacturer Narrative
Upon reception of the senor in-house, it was investigated, and the s1 complaint (unable to link sensor) was confirmed, as the identification of the sensor as well as measurements were not possible using both the qc rigs and a good working transmitter.The root cause of this failure is likely asic damage within the sensor, affecting its ability to communicate.As part of resolution, the rma was authorized to offer the user a sensor replacement.D9.Device available for evaluation? yes, 24 august 2021.H3.Device evaluated by manufacturer? yes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20875-7005
MDR Report Key12353969
MDR Text Key267672551
Report Number3009862700-2021-00103
Device Sequence Number1
Product Code QCD
UDI-Device Identifier00817491022349
UDI-Public00817491022349
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/20/2020
Device Model Number102096-67A
Device Catalogue NumberFG-4200-00-301
Device Lot NumberWP08273
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/26/2021
Initial Date FDA Received08/23/2021
Supplement Dates Manufacturer Received07/26/2021
Supplement Dates FDA Received03/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age29 YR
Patient SexFemale
-
-