Model Number 1616070
Device Problems Unsealed Device Packaging (1444); Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 10/2021).
 
Event Description
It was reported that prior to a procedure, the package was allegedly found to be already opened.There was no patient contact.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed powerport clearvue kit and outer packaging box was returned for evaluation.Gross visual evaluation was performed.The investigation is inconclusive for the reported unsealed device packaging issue, as the exact circumstances at the time of the reported event are unknown, and the reported event could not be reproduced in the lab.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 10/2021).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that prior to a port placement procedure, the package was allegedly found to be opened already.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT CLEARVUE SLIM IMPLANTABLE PORT
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key12357513
MDR Text Key267813067
Report Number3006260740-2021-03460
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741026461
UDI-Public(01)00801741026461
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1616070
Device Catalogue Number1616070
Device Lot NumberREES3810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/28/2021
Initial Date FDA Received08/24/2021
Supplement Dates Manufacturer Received11/08/2021
Supplement Dates FDA Received11/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-