B. BRAUN SURGICAL, S.A. DAFILON BLUE 5/0 (1) 45CM DS12; OTHER SUTURE Back to Search Results
Model Number C0932078
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed in the market (b)(4).There are no units in stock in b.Braun surgical's warehouse.Without closed samples and/or defective samples a proper analysis cannot be performed.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfils usp/ep and bbs requirement.Furthermore, needle attachment results conducted on samples before releasing the product were 0.71 kgf in average and 0.67 kgf in minimum and fulfilled the requirements of the european pharmacopoeia: 0.23 kgf in average and 0.11 kgf in minimum.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Event Description
It was reported an issue with dafilon suture.The client reported that they found the needle and thread separated, when opening the product packaging.There is no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 5/0 (1) 45CM DS12
Type of Device
OTHER SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key12358595
MDR Text Key267830870
Report Number3003639970-2021-00413
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K151165
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0932078
Device Catalogue NumberC0932078
Device Lot Number620344
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/29/2021
Initial Date FDA Received08/24/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-