Testing of actual/suspected device (10/213): a getinge field service engineer (fse) evaluated the iabp and was able to verify issue, the unit powered up right when it was plugged in and then a loud audible alarm.To fix the issue, the fse replaced the power management board and the unit is working now.Safety, and functionality checks completed per the service manual and passed to the factory specifications.Returned for clinical service upon completion.A supplemental report will be submitted upon completion of our investigation.
|