• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problems Device Alarm System (1012); Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device (10/213): a getinge field service engineer (fse) evaluated the iabp and was able to verify issue, the unit powered up right when it was plugged in and then a loud audible alarm.To fix the issue, the fse replaced the power management board and the unit is working now.Safety, and functionality checks completed per the service manual and passed to the factory specifications.Returned for clinical service upon completion.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) had a high pitch alarm.There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12362414
MDR Text Key267966608
Report Number2249723-2021-01895
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/02/2021
Initial Date FDA Received08/24/2021
Supplement Dates Manufacturer Received02/15/2024
Supplement Dates FDA Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-