• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. MODULAR NECK B 12/14 NECK TAPER USE WITH +0 HEADS ONLY; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. MODULAR NECK B 12/14 NECK TAPER USE WITH +0 HEADS ONLY; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: foreign country: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that when the sterile packaging of this product was opened, a substance such as a piece of vinyl, was found in the sterile packaging.This product was used for the patient.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual evaluation of the returned product found the product was returned opened.Presence of an unknown white debris was confirmed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was returned opened.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR NECK B 12/14 NECK TAPER USE WITH +0 HEADS ONLY
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12365608
MDR Text Key268067486
Report Number0001822565-2021-02397
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00889024137882
UDI-Public(01)00889024137882(17)310113(10)64905198
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K182678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00784802200
Device Lot Number64905198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/09/2021
Initial Date FDA Received08/25/2021
Supplement Dates Manufacturer Received11/16/2021
Supplement Dates FDA Received11/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-