• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Insufficient Cooling (1130)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon was duplicated and the cooling fan of the subject device was broken.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the preparation for use, the cooling fan of the lamp side of the subject device did not work.The event date was unknown.Other detailed information was not provided.There was no report of patient injury associated with the event.This device is an olympus asset.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) could not investigate the subject device, because the subject device was not returned to omsc.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Since the subject device was not returned to omsc, the exact cause was unknown.Approximately 8 years have passed since the subject device was manufactured.Omsc surmised that the reported phenomenon occurred since the cooling fan of the subject device was broken due to aging degradation, if it was used frequently.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12370004
MDR Text Key268215806
Report Number8010047-2021-10775
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCLV-S190
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/05/2021
Initial Date FDA Received08/26/2021
Supplement Dates Manufacturer Received09/06/2021
Supplement Dates FDA Received09/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-