Manufacturer's investigation conclusion: the reported event of the console stopping charging and switching to battery mode was confirmed.The centrimag 2nd generation primary console (serial #: (b)(6)) was returned for analysis to the service depot and the reported event was able to be duplicated.The console was opened, and damaged components were observed on the power supply.When ac power was supplied, the console would not operate.The console was further inspected, and an open fuse was found in the power entry module.The power supply and fuse were replaced, resolving the issue.The console was functionally tested and passed all tests.The power supply and fuse were returned to product performance engineering (ppe) for further analysis.The fuse was measured with a digital multimeter and was confirmed to be open.The power supply was inspected and there was no visible damage.The power supply was unable to be further troubleshot due to it causing damage to fuses.The root cause for the reported event was conclusively determined to be an issue with the 24v power supply, but a further root cause was unable to be conclusively determined.The device history records were reviewed for the centrimag 2nd generation primary console (serial #: (b)(6)) and the console was found to pass all manufacturing and qa specifications.The battery (serial #: (b)(6)) was found to have shipped with this console.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." no further information was provided.The manufacturer is closing the file on this event.
|