• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2021
Event Type  malfunction  
Manufacturer Narrative
The customer called in to report this event to olympus technical assistance center (tac) personnel.He noted that when the camera was disconnected from the cv-180, the blinking indicator lights from the control panel stop blinking and the color bar set appears on the display (as expected).The customer is planning to test the imager using other devices in attempt to run down the root cause of the light error.The device has not been returned for evaluation.If additional information becomes prior to the conclusion of the investigation, a supplemental report will be filed.
 
Event Description
The biomedical engineer reported that the loaned evis exera ii video system center began blinking status lights.The initial reporter went on to note that this was not discovered during a procedure, but during check out.No patient harm or consequence reported as a result of this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the manufacturer¿s investigation.A review of the device history record (dhr), as well as a historical trending analysis were conducted during this investigation.The review of the dhr did not find any abnormalities or anomalies identified during production.The device met all specifications upon release.Based on the results of the investigation as well as the manufactured date of the device, investigation believes fatigue was the cause of the reported event.More than 14 years have passed since the manufactured date of the device.It is likely that a deterioration of device components caused the leds to begin blinking.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12388215
MDR Text Key280777538
Report Number8010047-2021-10971
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/05/2021
Initial Date FDA Received08/30/2021
Supplement Dates Manufacturer Received10/06/2021
Supplement Dates FDA Received10/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-