• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Model Number AB2000
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2021
Event Type  malfunction  
Manufacturer Narrative
A root cause for the reported event is yet to be determined; investigation by the manufacturer is currently in-process.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during the aquablation procedure the aquabeam scope visualization fiber was broken and had to be replaced to continue the procedure through successful completion.The reported event caused a surgical procedural delay of over 40 minutes.There were no adverse health consequences with the patient because of this event.
 
Manufacturer Narrative
H10: additional manufacturer narrative: the aquabeam scope was returned for investigation.The reported aquabeam scope visualization fiber being broken could not be confirmed.The scope was returned in one (1) piece.Further investigation confirmed a visualization issue when looking through the eyepiece no image could be seen.The image was completely black indicating that the visualizatio fiber (fiber optic) was broken.A review of the device history record (dhr) for serial number (b)(6) and the lot number 2011 for the aquabeam scope were performed, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system and the aquabeam scope met all design and manufacturing specifications when released for distribution.A review of similar complaints for fiber optic breaks confirmed 17 other similar events have been reported to procept.The aquabeam robotic system user manual, um0101-00 rev.F, was reviewed and states the following: 11.2.5 sterile: aquabeam handpiece and aquabeam scope setup: verify the aquabeam scope visualization via the camera source port.Note: obtain a new aquabeam scope if proper visualization cannot be verified.The reported event of a broken scope was not confirmed; however, additional analysis confirmed a broken fiber optic on the returned scope, which caused no image.The root cause was unable to be conclusively determined.Procept has identified a trend in visualization issues with the aquabeam scope, which are currently being addressed through the corrective action and preventive action process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer Contact
doria esquivel
900 island drive
suite 101
redwood city, CA 94065-1494
6502327291
MDR Report Key12389297
MDR Text Key270824122
Report Number3012977056-2021-00066
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-PublicB614AB20001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAB2000
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/04/2021
Initial Date FDA Received08/30/2021
Supplement Dates Manufacturer Received01/12/2022
Supplement Dates FDA Received01/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-