H10: additional manufacturer narrative: the aquabeam scope was returned for investigation.The reported aquabeam scope visualization fiber being broken could not be confirmed.The scope was returned in one (1) piece.Further investigation confirmed a visualization issue when looking through the eyepiece no image could be seen.The image was completely black indicating that the visualizatio fiber (fiber optic) was broken.A review of the device history record (dhr) for serial number (b)(6) and the lot number 2011 for the aquabeam scope were performed, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system and the aquabeam scope met all design and manufacturing specifications when released for distribution.A review of similar complaints for fiber optic breaks confirmed 17 other similar events have been reported to procept.The aquabeam robotic system user manual, um0101-00 rev.F, was reviewed and states the following: 11.2.5 sterile: aquabeam handpiece and aquabeam scope setup: verify the aquabeam scope visualization via the camera source port.Note: obtain a new aquabeam scope if proper visualization cannot be verified.The reported event of a broken scope was not confirmed; however, additional analysis confirmed a broken fiber optic on the returned scope, which caused no image.The root cause was unable to be conclusively determined.Procept has identified a trend in visualization issues with the aquabeam scope, which are currently being addressed through the corrective action and preventive action process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|