RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+ |
Device Problem
Disconnection (1171)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/03/2021 |
Event Type
malfunction
|
Event Description
|
The customer reported "patient disconnect" alarm occurred.The field service engineer (fse) could not duplicate the issue.The service engineer (se) that visited onsite confirmed that when a circuit was removed, then the reported alarm sounded and diagonal line(s) indicating that the alarm was resolved quickly.However, the high-level alarm remained sounding.The field service engineer (fse) re-ran the unit under performance verification testing (pvt) and no anomaly was found.The unit was returned to service.The device was not in use on a patient.No patient or user harm was reported.
|
|