• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number NX1000-3
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Arrhythmia (1721)
Event Date 08/05/2021
Event Type  Death  
Manufacturer Narrative
The device was received for evaluation and successfully passed testing.Available log files were retrieved and analyzed which showed the cycler performed as intended and alarmed appropriately when the alarm conditions were present.A review of the device history record (dhr) was conducted which confirmed that the device met all quality criteria and manufacturing specifications prior to release.There was no indication of a device malfunction from the available information.The nxstage system one user guide supplement for solo home hemodialysis outlines risks, warnings, and requirements for the solo home hemodialysis patient.Udi: (b)(4).
 
Event Description
A report was received on 06 aug 2021 from the home therapy nurse (htn) of a (b)(6) male with a medical history including hypertension and end stage renal disease approved for performing solo home hemodialysis treatment, who stated the patient expired at an unspecified time during a home hemodialysis treatment on (b)(6) 2021.Additional information was received on 06 aug 2021 from the htn who stated the patient was found unresponsive at 7:40 by his son.911 was called and emergency services performed cardiopulmonary resuscitation (cpr) before the patient was pronounced at the scene.Per the htn and physician, the patent's death is due to a sudden cardiac event, unrelated to nxstage product or therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
MDR Report Key12413377
MDR Text Key269532220
Report Number3003464075-2021-00039
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K122051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberNX1000-3
Device Catalogue NumberCHRONIC HI-FLOW CYCLER
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/06/2021
Initial Date FDA Received09/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age57 YR
Patient Weight84
-
-