Other text: additional information added to h6 and h10.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found sample presented damage/broken on the breathing bag.Functional testing confirmed the issue; sample was rejected by the leak test equipment due to damage / broken on the breathing bags.The root cause of the reported issue, based on the investigation results, there is leakage due to pinholes on the raw material breathing bag provided by the supplier.Actions were taken to mitigate the reported issue: corrective actions was addressed and implemented through supply notification.
|