• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATUNE PRESSFIT STR STEM12X60MM; KNEE STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATUNE PRESSFIT STR STEM12X60MM; KNEE STEMS Back to Search Results
Model Number 1513-12-060
Device Problem Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2021
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Compass case report (b)(4): this patient has a left total attune knee done by surgeon.It is unknown when the knee was done.On xray, the femur has osteolysis and the tibia looks loose and has fallen into varus.The surgeon revised these components, and it turns out the femur and tibial tray were well fixed and difficult to remove.The patella was well fixed and retained.New implants were inserted.The tibia was prepped for an attune revision tray.The stem was overreamed by 1mm, the broach was fully seated.The real implant did not sit fully.It sat up 2mm.This seems to be a common event when using the attune revision system.All the trials sat perfectly, and the real implant would not seat fully.The pressfit is too much.No delay in surgery.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received indicated that the revision was due to malalignment.The patient had pain.The tray sitting up required the use of a thinner insert.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received, will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers, the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: a review of the device manufacturing records (dhr) was performed as part of this investigation.A search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude, that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUNE PRESSFIT STR STEM12X60MM
Type of Device
KNEE STEMS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key12422768
MDR Text Key271502501
Report Number1818910-2021-19323
Device Sequence Number1
Product Code JWH
UDI-Device Identifier10603295043546
UDI-Public10603295043546
Combination Product (y/n)N
PMA/PMN Number
P830055
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 02/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1513-12-060
Device Catalogue Number151312060
Device Lot NumberJ81N00
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/08/2021
Initial Date FDA Received09/03/2021
Supplement Dates Manufacturer Received09/03/2021
10/27/2021
Supplement Dates FDA Received09/27/2021
10/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
5 LEFT PS ATTUNE FEMUR; 5X12MM ATTUNE PS RP INSERT; ATUN TIB SLV M/L 37MM FULL POR; ATUNE PRESSFIT STR STEM12X60MM; ATUNE REV RP TIB BASE SZ 3 CEM; UNK ATTUNE KNEE PATELLA; UNK KNEE TIBIAL TRAY ATTUNE RP
-
-