Evaluation of the returned catrx confirmed that the catheter was kinked.The observed kinks on the device likely occurred during use and handling.If the catrx is mishandled at an extreme angle during use, damage such as a kink may occur.Further evaluation revealed that the catrx was fractured and had bends in the catheter shaft.This damage was incidental to the reported complaint.The fracture likely occurred due to manipulation of the kink.However, the root cause of the bends could not be determined.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|