RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+ |
Device Problem
Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/11/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that when the customer started up the ventilator from a state of standby before use on a patient, a "check vent: backup alarm failed" alarm occurred.There was no patient involvement.
|
|
Manufacturer Narrative
|
The service provider (sp) inspected the device and confirmed the reported issue.The sp replaced the central processing unit (cpu) to address the reported issue.The unit was checked overall, run-in tested, cleaned, functionally tested, and no abnormality was confirmed.
|
|
Manufacturer Narrative
|
Upon further investigation it was reported that this event occurred outside of clinical use (testing unrelated to set up for patient use).Therefore, this complaint no longer meets reportability requirements.
|
|