• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN SHOULDER METAGLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN SHOULDER METAGLENE Back to Search Results
Catalog Number UNK SHOULDER METAGLENE
Device Problem Osseointegration Problem (3003)
Patient Problems Sepsis (2067); Inadequate Osseointegration (2646)
Event Date 08/24/2020
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
An article/literature was received entitled " stable glenoid component of reverse total shoulder arthroplasty at 2 years as measured with model-based radiostereometric analysis (rsa)".Literature article entitled "stable glenoid component of reverse total shoulder arthroplasty at 2 years as measured with model-based radiostereometric analysis (rsa)" written by alexander nilsskog fraser, berte bøe, tore fjalestad, jan erik madsen and stephan m röhrl.Published by acta orthopaedica published online/accepted by publisher 1 jul 2021 was reviewed.The article's purpose was to follow 20 patients and assess stability of the glenoid component in reverse tsa, using model-based rsa.The 20 patients received primary treatment with a reversed tsa using a delta xtend shoulder.Specific patient identifiers were not given in the article.X-ray and ct images can be found on pages 3 and 4 of the literature article.Depuy products with known adverse event(s): metaglene x 2, locking screw x 4, glenosphere, humeral cup, humeral epiphysis, humeral stem.Adverse events: migration of the metaglene with no indicated treatment.Humeral cup notching with no indicated treatment.Septic loosening with revision.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed, and evidence of shoulder metaglenes and shoulder screws migration was found.No evidence of implant loosening was noted.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN SHOULDER METAGLENE
Type of Device
SHOULDER METAGLENE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12437073
MDR Text Key270841404
Report Number1818910-2021-19506
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Type of Report Initial,Followup
Report Date 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK SHOULDER METAGLENE
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/17/2021
Initial Date FDA Received09/07/2021
Supplement Dates Manufacturer Received09/16/2021
Supplement Dates FDA Received09/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-