• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/12/2021
Event Type  malfunction  
Event Description
A user facility reported to olympus that, the evis exera ii video system center brightness adjustment does not work.There was no harm, infection or user injury reported due to the event.
 
Manufacturer Narrative
The device was returned for evaluation.The evaluation was unable to confirm the reported event of the auto iris and the auto brightness was not working.The video processor was tested with a light source and video images functioned normally.The auto adjustment was also working okay.The device was inspected and passed olympus functional standards.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Manufacturer Narrative
The supplemental report is submitted to provide the result of the legal manufacturer¿s investigation.The device history record for the subject device was reviewed and it was verified that the device was manufactured in accordance with documented specifications.The legal manufacturer performed an investigation.A conclusive root cause was not identified.The investigation determined that there was no defect in the equipment, but there was a defect with the light source device and the light source cable itself, or the connection of the light source cable was not appropriate.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states: "properly and securely connect all cables.Otherwise, equipment damage or malfunction can result.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12442515
MDR Text Key280754048
Report Number8010047-2021-11444
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/16/2021
Initial Date FDA Received09/08/2021
Supplement Dates Manufacturer Received10/14/2021
Supplement Dates FDA Received10/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-