One single dpt - vamp plus kit with saline bag was returned for examination.Both red and blue dpt sensors zeroed and sensed pressure on the laboratory patient monitor.No error message was noticed on the monitor.Pressure did not show any drift during output drift testing and met specifications.Electrical testing showed that both input and output impedances were within specifications.The zero-offset also met specifications.No leakage or occlusion was detected from the kit during pressure test.No visible damage was observed from the kit.A device history record review was completed and documented that the device met all specifications upon distribution.As per ifu, poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patients clinical manifestations.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
As reported, as a control study, after a coronary angiography/ percutaneous coronary intervention on a patient in the cardiology lab with this disposable pressure transducer (dpt) with vamp, the blood pressure measured 174/75 mmhg while that of the cardiologist using acist measured 164/74 mmhg.In order to take these measures, both transducers were connected to the sheat that was previously attached to the patient by the cardiologist.No medication was given based on these values.There was no allegation of patient injury.
|