MEDTRONIC MEXICO S. DE R.L. DE CV BIO-MEDICUS STANDARD INSERTION KIT; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 96553 |
Device Problem
Device Markings/Labelling Problem (2911)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/08/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Reason for return was confirmed.Visual inspection on the peel pouch label shows the guidewire is to be 0.025 inches in diameter.The guidewire was measured using a micrometer to be 0.037 inches in diameter.The guidewire appears to have some kinks.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of a 96553 bio-medicus nextgen insertion kit, the customer reported that the guide wire could not be inserted in the sheath.The size was large and it was difficult to pass through, so the product was not used.Additional information received: when replacing the guide wire the customer needed to puncture another site and in order to stop any bleeding a suture was required.The procedure was extended by approximately 15 mins.There was no adverse effect to the patient.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|