Catalog Number 07671687003
Device Problem Low Readings (2460)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/30/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter occupation is a lay user/patient.The meter and test strips were requested for investigation.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.There were 9 test strips of lot 507723-11 tested for the investigation.Testing results (qc range = 2.5 - 3.1 inr): qc 1: 2.8 inr.Qc 2: 2.8 inr.Qc 3: 2.8 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.Product labeling states, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." the investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
There was an allegation of discrepant inr results for one patient tested with a coaguchek xs meter serial number (b)(4) compared to a laboratory test with an unknown reagent.The result from the meter on (b)(6) 2021 between 8:00 am and 9:00 am was 3.3 inr.The result from the laboratory on (b)(6) 2021 at 11:34 am was 4.43 inr.The result from the meter on (b)(6) 2021 between 8:00 am and 9:00 am was 2.4 inr.The result from the laboratory on (b)(6) 2021 at 11:16 am was 3.39 inr.The patient¿s therapeutic range is 2.5- 3.5 inr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key12453126
MDR Text Key272781631
Report Number1823260-2021-02653
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number07671687003
Device Lot Number50772311
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2021
Initial Date Manufacturer Received 08/19/2021
Initial Date FDA Received09/10/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age82 YR
-
-