THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
|
Back to Search Results |
|
Model Number 102956 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/16/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
|
|
Event Description
|
It was reported that the centrimag motor had a stuck set screw.The customer site will return the centrimag motor for analysis and repair.
|
|
Manufacturer Narrative
|
Manufacturer's investigation conclusion: the reported event of a damaged set screw was confirmed.The centrimag motor (serial #: (b)(6)) was returned for analysis and was evaluated and tested.The reported event was visually verified.The locking screw was jammed and unable to rotate.The customer never responded to the purchase order request to fulfil the repair.The unit was sent back to the customer unrepaired.The root cause for the reported event was unable to be conclusively determined through this analysis.The device history records were reviewed for the centrimag motor and the motor was found to pass all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." no further information provided.The manufacturer is closing the file on this event.
|
|
Search Alerts/Recalls
|
|
|