Catalog Number 2C8541 |
Device Problems
Filling Problem (1233); Material Twisted/Bent (2981); No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a clearlink system continu-flo solution set under infused.This occurred during infusion with calcium chloride.It was stated that the new bag hung only had about 100 ml gone where almost 500 ml should have infused.When the tubing was taken out of the pump it was noted it was completely flat and occluded.It was further reported that the unspecified pump never alarmed and stated the volume had infused.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional test including pressure test and clear passage test was performed with no defects observed.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|