MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
|
Back to Search Results |
|
Model Number BI70000028230 |
Device Problems
Failure to Run on Battery (1466); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr: a manufacturer rep went to the site to test the system.The system passed the system checkout after parts were replaced.Concomitant medical products: other relevant device(s) are: product id: b i31000478 ,product id: bi31000479, product id: bi71000524, product id: bi71000581, product id: bi71000126, product id: bi71000125, product id: bi71000514, product id: bi71000480.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the image acquisition system (ias) shut down automatically during post operative inspection. it was found that voltage of motor battery was low. also, it was reported that mvs sometimes restarted by itself. no patient injury was involved.
|
|
Manufacturer Narrative
|
H2: additional information: d10: bi71000480, lot no:s719100399; bi71000477, lotno:s1311ljx; bi71000486, lotno:s1248orc h3: the board and uninterruptible power supply (ups) were returned to the manufacturer for analysis.Analysis found that the reported issue could not be confirmed with either component.No failure was found while under testing.(b01, c19, d15) the motor battery charger board was returned to the manufacturer for analysis.Analysis found that the reported issue could not be confirmed; however, the board was electrically overstressed (b01, c02, d02).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: the mobile view station (mvs) power board (part: bi71000514, return part: bi71000514) was returned to the manufacturer for analysis.Analysis found that the board arrived with two missing fuses and it was not possible to test further.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|