• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2R
Device Problem Material Perforation (2205)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2021
Event Type  malfunction  
Event Description
The customer originally returned the olympus uretero-reno videoscope due to the device displaying a 'bad image'.The initial reporter did confirm that this 'bad image' did not occur during a procedure.No patient contact.Upon device evaluation, it was discovered that the distal end of the device was damaged and the metal was exposed.This report is being submitted to capture the damaged end with the exposed metal.
 
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user's report was unable to be confirmed.The scope was presenting a good image with no stains noted.However, the scope was broken on the bending section skeleton, puncturing the distal end sheath and causing a leak to occur.If additional information becomes available following device evaluation, a supplemental report will be filed.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Based on the results of the investigation, it is likely that the user manipulated the device so as to apply excessive stress to the bending section, which damaged the bending tube.User can properly detect the suggested event by handling device according to the following ifu description: ¿precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk.Inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.¿ user can reduce / prevent occurrence of the suggested event by handling device according to the following ifu description: ¿do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.¿ olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12461087
MDR Text Key280811152
Report Number8010047-2021-11629
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343612
UDI-Public04953170343612
Combination Product (y/n)N
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/24/2021
Initial Date FDA Received09/13/2021
Supplement Dates Manufacturer Received10/25/2021
Supplement Dates FDA Received10/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-