The lot # was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
|
Titan otr pump, cylinders 1 and 2, and reservoir were received for evaluation.A separation within abrasion was noted on the inlet tube of the pump.This was a site of leakage.Partial separations within abrasion were also noted on the inlet tube near the separation site.These were not sites of leakage.Partial separations within abrasion were noted on the exhaust tube of cylinder 2.This was not a site of leakage.No functional abnormalities were noted with cylinders 1 and 2.A partial separation was noted on the inlet tube near the tube/strain relief junction of the reservoir.This was not a site of leakage.No functional abnormalities were noted with the reservoir.Based on recreation of the position of the tubes according to the abrasion pattern, this demonstrates that the exhaust tube of cylinder 2 and inlet tube were overlapping while in vivo.This positioning, in combination with device usage over time, may contribute to sufficient stress(s) to separate the exhaust tubing of cylinder 2.A separation of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.D4 lot: 1780100.
|