The device was returned to olympus for evaluation and the issue was not confirmed.The air/water supply passed the capacity test.All switches were scratched.The distal end plastic cover had dents and the pin was old.The objective lens and light guide lens had old glue.The nozzle pin was old.There was low angulation and scratches at the light guide lens tube, connector, and s-cover.The labeling was peeling and illegible.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, it is likely that a part of peripheral device which was used in combination with the subject scope entered the air/water channel.It is also likely that foreign material left in the air/water channel clogged the nozzle by inadequate reprocessing on the channel.A definitive root cause cannot be identified.Ifu (reprocessing manual) specifies about inspection prior to use for the distal end as below: ¿3.3 inspection of the endoscope 8.Inspect the air/water nozzle at the distal end of the endoscope¿s insertion section for abnormal swelling, bulges, dents, or other irregularities.¿ ifu (operation manual) specifies about inspection prior to use for the water feeding function as below: ¿3.8 inspection of the endoscopic system inspection of the objective lens cleaning function depress the valve.Observe the endoscopic image and confirm that water flows on the entire objective lens.After checking the water feeding function and while observing the endoscopic image, feed air by covering the hole in the air/water valve with your finger.Confirm that the emitted air removes the remaining water from the objective lens and clears the endoscopic image.¿ ifu (reprocessing manual) specifies about nozzle reprocessing to be taken immediately after procedure e as below: ¿5.3 precleaning the endoscope and accessories flush the air/water channel with water and air: caution to prevent clogging of the air/water nozzle of the endoscope, flush water into the air channel of the endoscope, using the aw channel cleaning adapter (mh-948) after each patient procedure.¿ olympus will continue to monitor the field performance of this device.
|