Model Number 102956
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2021
Event Type  malfunction  
Manufacturer Narrative
No patient involved.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event Description
It was reported that the centrimag motor had a bent locking screw.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of the centrimag motor¿s locking screw being damaged was confirmed, as the returned centrimag motor (serial number (b)(6)) was observed to have a damaged locking screw upon arrival.The head of the screw was observed to be broken off, and the screw was unable to rotate.The motor was received by the service depot and its locking mechanism was replaced with a new component.Then, the motor was functionally tested alongside known working test equipment and was found to perform as intended.The serviced and repaired motor was returned to the customer site after passing all tests per procedure.The root cause of the reported event was unable to be conclusively determined through this analysis.Review of the device history record for centrimag motor, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual provides information regarding emergencies/troubleshooting in section 10.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console, and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.Centrimag motor ifu instructs the user to inspect the centrimag motor, cable, console connector, and locking mechanism for any damage prior to use.If any component is damaged, do not use the centrimag motor.This document states that if the unit fails to operate according to the motor specifications or a console diagnostic error indicates a centrimag motor malfunction, it should be returned.Additionally, this document instructs the user to always have a spare centrimag motor and back-up equipment available.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key12466999
MDR Text Key271254347
Report Number3003306248-2021-04034
Device Sequence Number1
Product Code KFM
UDI-Device Identifier07640135140078
UDI-Public07640135140078
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020271
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102956
Device Catalogue Number102956
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/24/2021
Initial Date FDA Received09/14/2021
Supplement Dates Manufacturer Received01/14/2022
Supplement Dates FDA Received01/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-