FORESIGHT ELITE OXIMETER; FSEO2004 Back to Search Results
Model Number FSEO2004
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2021
Event Type  malfunction  
Manufacturer Narrative
The product has not been returned for evaluation.When it arrives and the product evaluation has been completed a supplemental report will be submitted.The device history record review was not completed but will be provided after it has been finished.The udi number is not available.
 
Event Description
It was reported that the fseo2004 started giving numbers that are outside the simulator calibrated range.When the simulator is connected the readings should be within 10 percent of 65.The readings are either high or low and thus not calibrated.This was before use.There is no patient injury.Patient demographics were requested but unavailable.
 
Manufacturer Narrative
One foresight elite oximeter was received for product evaluation.A visual inspection found that both of the sensor cables strain reliefs on the module ends are damaged.The suspect unit was connected to a known good working system.Using simulators, the sto2 readings of all 4 channels showed 65 percent plus or minus 5 percent and remained steady even while moving the cables.No error messages were observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORESIGHT ELITE OXIMETER
Type of Device
FSEO2004
MDR Report Key12468849
MDR Text Key271314726
Report Number2015691-2021-05171
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
PMA/PMN Number
K142749
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFSEO2004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/18/2021
Initial Date FDA Received09/14/2021
Supplement Dates Manufacturer Received10/13/2021
Supplement Dates FDA Received10/21/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-