OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2021
Event Type  malfunction  
Manufacturer Narrative
The device was returned to the service center for evaluation.The reported issue was confirmed as the main switch was damaged, difficult to activate, and requires upgrade to latest switch version.Additionally, the scope connector socket tab is not protecting socket pins.The investigation is ongoing; therefore, the root cause of the reported issue/malfunction cannot be determined at this time.However, if additional information becomes available a follow up medical device report will be supplemented accordingly.
 
Event Description
The technical support engineer was informed by the biomedical equipment technician at the user facility that during preparation for use, the evis exera iii xenon light source reportedly was found to have a broken power button malfunction as the power switch cannot be pushed.No patient involvement or harm was reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the event was due to wear.The power switch was stuck because the operating force and the number of times the power switch was applied exceeded the assumption and design.There was a design change implemented to the power switch to improve the resistance to damage of the power switch.Countermeasure products have been shipped after november 26, 2013.The subject device was manufactured before the design change.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12471462
MDR Text Key280874316
Report Number8010047-2021-11705
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/26/2021
Initial Date FDA Received09/14/2021
Supplement Dates Manufacturer Received11/04/2021
Supplement Dates FDA Received11/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-