SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR TIP; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71363077 |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/24/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference (b)(4).
|
|
Event Description
|
It was reported that, during a thr surgery the tip of a r3 offset impactor tip was split when impacting the cup.Surgery was finished with a smith and nephew back up device, without any surgical delay.All pieces were retrieved.Patient was not injured as consequence of this problem.
|
|
Manufacturer Narrative
|
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|