Catalog Number 8065751765
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported infusion line and the gas line were separated during the vitrectomy procedure.The procedure was completed.There was no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.One unused combined pak was returned.Visual inspection confirmed that the auto stopcock was broken.The top and base surface profile was inspected and demonstrated an inconsistent shear weld at the interfaces.The root cause of the customer's complaint is due to a weak weld between the top and base of the auto infusion valve housing.This complaint has been reviewed and it is determined that no further actions will be pursued at this time.After investigation of this complaint and analysis of complaints of this nature confirm no unfavorable trend for this event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis by the product team.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12479745
MDR Text Key271652628
Report Number1644019-2021-00565
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657517657
UDI-Public00380657517657
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number8065751765
Device Lot Number2458974H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/08/2021
Initial Date FDA Received09/16/2021
Supplement Dates Manufacturer Received01/06/2022
Supplement Dates FDA Received01/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-