• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problems Defective Device (2588); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was received and evaluated at olympus (b)(6) service center.Inspection found the device scope detection does not work.Scope detection skips.Furthermore, the device base plate and lid were both found corroded.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event Description
As reported, during device service inspection it was found that the device scope detection does not work.There is no patient involvement associated on this reported event.No user injury reported.Prior to the event reported, the device was used for an endoscopy examination which was completed successfully with the same device.There was no extension of the procedure and no injury to the patient.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the investigation, and it is likely that the parts of the slide switch were worn out due to repeated use or accidentally broken down.If the corners of the slide switch parts become rounded, it is probable that the switch became difficult to push in when the scope was inserted, making it impossible to detect it.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12485557
MDR Text Key274981777
Report Number8010047-2021-11853
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/30/2021
Initial Date FDA Received09/16/2021
Supplement Dates Manufacturer Received11/05/2021
Supplement Dates FDA Received11/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-