• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Fail-Safe Did Not Operate (4046)
Patient Problem Laceration(s) (1946)
Event Date 07/27/2021
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the perforator failed to disengage when drilling the first hole during a left parieto-occipital craniotomy for left occipital lobectomy in a (b)(6) year-old child.The bit/tip of perforator suddenly entered into intracranial space and tore the dura with wide opening/laceration of the superior longitudinal sinus with significant blood loss.The situation could be controlled and the sinus was sutured.The initial planned surgery could not be performed as medical staff would like to check health state of the patient.Medical staff did postoperative ct angiography to confirm the absence of stenosis or sinus obstruction.The initial planned surgery has to be rescheduled / patient had to stay at the hospital longer period.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10 the perforator was returned for evaluation: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the perforator unit was inspected using the unaided eye.Unit was lightly soiled but passed visual inspection and showed no observable anomalies.Ifu testing was performed with no observed anomalies.Functional was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
MDR Report Key12485703
MDR Text Key271890423
Report Number3014334038-2021-00193
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 10/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ64N53
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2021
Initial Date Manufacturer Received 09/01/2021
Initial Date FDA Received09/16/2021
Supplement Dates Manufacturer Received09/29/2021
Supplement Dates FDA Received10/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age3 YR
-
-