U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,INTL; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,INTL; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-UC-3013-55
Device Problems Electrical /Electronic Property Problem (1198); Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device (10/213): a getinge field service engineer (fse) evaluated the iabp unit for the reported "error# 35" and the found that the main board to the motor controller was defective.To fix the issue, the fse replaced the main board to motor controller pcb cable (part no- 0012-00-0759-02), and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.(b)(6).
 
Event Description
It was reported that during a routine check the cs100 intra-aortic balloon pump (iabp) was showing an electrical test failure code #53 intermittently.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Additional information: event site postal code: (b)(6), event site state: (b)(6).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,INTL
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12488277
MDR Text Key271908936
Report Number2249723-2021-02116
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107875
UDI-Public10607567107875
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-UC-3013-55
Device Catalogue Number0998-UC-3013-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/26/2021
Initial Date FDA Received09/17/2021
Supplement Dates Manufacturer Received02/16/2024
Supplement Dates FDA Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/06/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-