Incidental findings: battery cable damage, s1 alarm manufacturer's investigation conclusion: the reported event of a b1 alarm was confirmed.The centrimag 2nd generation primary console (serial #: (b)(6)) was returned for analysis and a log file was downloaded for review with events spanning approximately 12 days ((b)(6) 2021 per time stamp).Events occurring on (b)(6) 2021 took place during lab testing at abbott.Beginning on (b)(6) 2021 at 16:38, battery module fail: b1 alarms began to activate.These alarms were able to be muted but not cleared.They continued to activate until (b)(6) 2021 at 10:39.There were no other notable alarms active in the log file.Pump operation was not affected.The centrimag 2nd generation primary console was returned for analysis to the european distribution center (edc) and the reported event was able to be reproduced.The sps printed circuit board (pcb) was replaced, resolving the issue.Functional testing and preventative maintenance were performed.The pcb was forwarded to product performance engineering (ppe) for further analysis.Upon further analysis, the returned sps pcb was in unremarkable condition.The pcb was connected to a test fixture and motor, monitor, flow probe, and mock loop.The reported b1 alarm was able to be reproduced.A digital multimeter was used to probe the pcb and compare its values to a functioning test pcb.A few components were suspected to be related to the reported event and were subsequently replaced.Replacement of these components did not resolve the issue.An additional component, d502, was also suspected to be related to the event; however, due to its location on the pcb, it was unable to be removed and replaced.D502 is responsible for esd protection of the battery and follows in the circuit where the battery is connected.Additional provided information communicated on 21sep2021 stated that there was no harm to the patient who was going on ecmo.The root cause for the reported event was unable to be conclusively determined through this analysis; however, it is suspected that an issue component d502 caused the reported event.The device history records were reviewed for the centrimag 2nd generation primary console and the console was found to pass all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual table 13 entitled ¿console alarms & alerts¿ addresses how to properly interpret and troubleshoot all system alarms including s1 and b1 alarms.No further information was provided.The manufacturer is closing the file on this event.
|