ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS TAQSCREEN MPX TEST V2; HUMAN IMMUNODEFICIENCY VIRUS TYPE 1 (HIV-1) GROUP M AND O RNA, HIV-2 RNA, HCV
|
Back to Search Results |
|
Catalog Number 05969484190 |
Device Problem
False Negative Result (1225)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/18/2021 |
Event Type
malfunction
|
Event Description
|
A customer from (b)(6) alleged the generation of discrepant results compared to the serology result when using the cobas taqscreen mpx test, v2.0, lot h02619.A donation generated a non-reactive result in pool of 6 and the serology result was hcv reactive.Resolution testing in pool of 1 generated a hbv reactive result.Repeat of of the resolution testing in pool of 1 generated a non-reactive result.No harm was alleged.An investigation is on-going.
|
|
Manufacturer Narrative
|
An investigation is on-going.A supplemental report will be filed upon completion of the investigation.Facility name was truncated due to characters limits.The complete facility name is (b)(6).(b)(4).
|
|
Manufacturer Narrative
|
An investigation of the reagent kit lot h02619 did not identify any product issue.The alleged sample were returned for investigation.The returned sample was tested in triplicate, using a quantitative hcv test (cobas® hcv, processing volume 650ul) and the sample generated a hcv target not detected result for all 3 replicates.A possible cause for the hbv and hcv reactive results at the customer site could be a laboratory contamination; however, the customer stated it was unlikely as they did not know of any potential source of contamination in their lab.An occult hbv infection cannot be excluded but as they are rare it does seem unlikely.The customer further stated that they suspected potential false positive serology results as an explanation for the non-reactive nat results.(b)(4).
|
|