MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DENALI JUGULAR SYSTEM; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DENALI JUGULAR SYSTEM; VENA CAVA FILTER Back to Search Results
Catalog Number DL950J
Device Problems Difficult to Insert (1316); Failure to Advance (2524); Material Deformation (2976); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2021
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the denali jugular system that are cleared in the us.The pro code and 510 k number for the denali jugular system are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 02/2024).
 
Event Description
It was reported that during vena cava filter placement via the internal jugular vein, the vena cava filter was allegedly unable to insert through the sheath and the filter was allegedly deformed.It was further reported that another device was used to complete procedure.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the denali jugular system that are cleared in the us.The pro code and 510 k number for the denali jugular system are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one denali filter kit was returned for evaluation.Following components were received: dilator, pusher catheter, introducer sheath, filter storage tube, touhy-borst adapter and filter.Filter was noted to be extending out of the storage tube with two of the legs crossed and damaged.No skiving or bowing noticed to the storage tube, no damage to the introducer sheath or dilator.Based on the findings, the investigation is confirmed for the reported failure to advance issue as the filter was noted to be extending out of the storage tube.However, the investigation is inconclusive for the reported activation failure including expansion failure as the reported event cannot be reconfirmed.A definitive root cause for the reported failure to advance and activation failure including expansion failure issues could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2024).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during vena cava filter placement via the internal jugular vein, the vena cava filter was allegedly unable to advance through the sheath.It was further reported that the filter was allegedly deformed.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI JUGULAR SYSTEM
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key12493974
MDR Text Key272114428
Report Number2020394-2021-01754
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00801741040818
UDI-Public(01)00801741040818
Combination Product (y/n)N
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberDL950J
Device Lot NumberGFFQ0658
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/15/2021
Initial Date Manufacturer Received 08/23/2021
Initial Date FDA Received09/20/2021
Supplement Dates Manufacturer Received10/06/2021
Supplement Dates FDA Received10/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-