Model Number 5833690 |
Device Problems
Fluid/Blood Leak (1250); Expulsion (2933); Material Protrusion/Extrusion (2979)
|
Patient Problems
Extravasation (1842); Fever (1858); Swelling/ Edema (4577)
|
Event Date 05/19/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 05/2021).
|
|
Event Description
|
It was reported that approximately five months and nine days post dialysis catheter placement in the right internal jugular vein, the catheter allegedly leaked and the patient experienced redness and swelling.It was further reported that the patient had fever and the polyester sleeve of hemodialysis catheter allegedly protruded, which was fixed with suture needle.The current status of the patient was unknown.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b2, d4 (expiry date: 05/2021), g3, h6 (patient, method).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that approximately five months and nine days post dialysis catheter placement in the right internal jugular vein, the polyester sleeve of catheter allegedly protruded.It was further reported that it was fixed with suture needle.Reportedly patient experienced fever, redness, extravasation and swelling.The current status of the patient was unknown.
|
|
Search Alerts/Recalls
|
|
|