ACESSA HEALTH INC. ACESSA PROVU HANDPIECE; COAGULATOR, LAPAROSCOPIC, UNIPOLAR (AND ACCESSORIES)
|
Back to Search Results |
|
Model Number 7300 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/26/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that on (b)(6) 2021, during an acessa procedure, the handpiece failed and could not perform the ablation.The patient had to be rescheduled for another procedure.No other information is available.
|
|
Manufacturer Narrative
|
Lot and serial number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
|
Search Alerts/Recalls
|
|
|