• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANATOMAGE INC. ANATOMAGE GUIDE; SURGICAL GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANATOMAGE INC. ANATOMAGE GUIDE; SURGICAL GUIDE Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Perforation (2001)
Event Date 01/20/2021
Event Type  Injury  
Manufacturer Narrative
According to our investigation, all production processes were properly followed.The trajectory deviation may have been due to the sleeve shifting due to force from the handle, and/or complications due to patient morphology.
 
Event Description
The guide was used for implant surgery.The doctor performed the initial osteotomy and observed that the trajectory appeared much more buccal than in the final plan.She placed a directional pin and took a post-op scan, which showed that the drill had perforated through the buccal plate.Surgery was aborted and the site was grafted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANATOMAGE GUIDE
Type of Device
SURGICAL GUIDE
Manufacturer (Section D)
ANATOMAGE INC.
3350 scott blvd. bldg 29
santa clara CA 95054
Manufacturer (Section G)
ANATOMAGE INC.
3350 scott blvd bldg 29
santa clara CA 95054
Manufacturer Contact
meera kler
3350 scott blvd. bldg 29
santa clara, CA 95054
4088851474
MDR Report Key12503022
MDR Text Key272443245
Report Number3008272529-2021-00019
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/30/2021
Initial Date FDA Received09/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age41 YR
-
-