• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71335754
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event Description
It was reported that, during a thr, a r3 20 deg xlpe acet lnr 36mm x 54mm had a black speck inside the plastic bag.This was noticed before use in patient.Surgery was performed, without any delay, with a smith and nephew back-up device instead.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
Section: h3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.The device's packaging has a black spec within it.However, this spec is small enough to comply with the device's specifications.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.Some potential probable causes could be but are not limited to packaging or manufacturing procedural/user error.Based on this investigation, the need for corrective action is not indicated.A field action review concluded that there are no actions related to this part number and failure mode.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 36MM X 54MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12509443
MDR Text Key272568716
Report Number1020279-2021-07033
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010598349
UDI-Public03596010598349
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71335754
Device Catalogue Number71335754
Device Lot Number21BM13936
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/31/2021
Initial Date FDA Received09/21/2021
Supplement Dates Manufacturer Received12/10/2021
Supplement Dates FDA Received12/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-