• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation but was returned to olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon was not duplicated.Also, sorc found that there were pinholes on the bending rubber and inside the instrument channel.Omsc reviewed the manufacture history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However based upon the information from sorc, omsc surmised that the reported phenomenon was attributed to the corrosion inside the subject device due to the water ingress from the pinhole of the bending rubber and/or the leakage point of the instrument channel.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during the preparation for use, it was found that brown liquid came out from the instrument channel of the subject device.The occurrence date of the event is unknown.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12511466
MDR Text Key272707989
Report Number8010047-2021-12120
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/27/2021
Initial Date FDA Received09/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-