CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE
|
Back to Search Results |
|
Model Number DBEC-125 |
Device Problems
Crack (1135); Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device analysis conclusion: the oad was returned to csi for analysis.Visual examination confirmed the saline line was cracked at the proximal end of the y-adaptor luer connector.The exact cause of the damage could not be determined.When tested, fluid leaked from the damaged area; however, no air was observed in the saline line.The oad functioned as intended during functional testing except for the observed fluid leak.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|
|
Event Description
|
Prior to insertion into the patient's body, a crack was observed in the y-adaptor saline tubing of the diamondback coronary orbital atherectomy device (oad).Fluid was leaking from the area, and air was drawn into the system.The oad was removed, and a second device was used to complete the procedure.
|
|