CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM; CORONARY ATHERECTOMY DEVICE Back to Search Results
Model Number DBEC-125
Device Problems Crack (1135); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2021
Event Type  malfunction  
Manufacturer Narrative
Device analysis conclusion: the oad was returned to csi for analysis.Visual examination confirmed the saline line was cracked at the proximal end of the y-adaptor luer connector.The exact cause of the damage could not be determined.When tested, fluid leaked from the damaged area; however, no air was observed in the saline line.The oad functioned as intended during functional testing except for the observed fluid leak.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Event Description
Prior to insertion into the patient's body, a crack was observed in the y-adaptor saline tubing of the diamondback coronary orbital atherectomy device (oad).Fluid was leaking from the area, and air was drawn into the system.The oad was removed, and a second device was used to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM
Type of Device
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
morgan hill
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key12512600
MDR Text Key272690307
Report Number3004742232-2021-00329
Device Sequence Number1
Product Code MCX
UDI-Device Identifier10850026568087
UDI-Public(01)10850026568087(17)230531(10)380104-1
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberDBEC-125
Device Catalogue Number7-10060-09
Device Lot Number380104-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/25/2021
Initial Date FDA Received09/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
-
-