TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AHCT |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/31/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The srt replaced the aux pcba.The monitor successfully powered up and passed the self test.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The service repair technician (srt) reported that during testing of the device at the service center, the blood parameter monitor had a heat damaged capacitor at the c107 position on the auxiliary (aux) printed circuit board assembly (pcba) causing the monitor to fail power up.There was no patient involvement.
|
|