SMITH & NEPHEW, INC. UNKN REDAPT ACET CUP; RETRACTOR Back to Search Results
Catalog Number UNKN01100108
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Joint Dislocation (2374)
Event Date 07/27/2021
Event Type  Injury  
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, it is noted within the attachments that no clinical/medical information is available.Smith and nephew has not received the explanted device and/or adequate materials to fully evaluate the complaint.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Event Description
It was reported that after a tha was performed on an unknown date, the patient experienced hip dislocation.A revision surgery was performed on an (b)(6) 2021.During the surgery the modular neck and the acetabular shell were explanted and replace by a new ones.Additionally information about the revision surgery is unknown.The current patient status is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN REDAPT ACET CUP
Type of Device
RETRACTOR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12522194
MDR Text Key273086039
Report Number1020279-2021-07097
Device Sequence Number1
Product Code GAD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKN01100108
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/22/2021
Initial Date FDA Received09/23/2021
Supplement Dates Manufacturer Received09/22/2021
Supplement Dates FDA Received09/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-